Related Experiment Video
Updated: Jun 28, 2026

Clean Sampling and Analysis of River and Estuarine Waters for Trace Metal Studies
Published on: July 1, 2016
Sequential extraction of U, Th, Ce, La and some heavy metals in sediments from Ortigas river, Spain
R Martin1, D M Sanchez, A M Gutierrez
1Unidad de Quimica Analitica, Programa de Quimica, C.I.E.M.A.T., Av. Complutense 22, E-28040 Madrid, Spain.
Abstract:
The extractable contents of U, Th, Ce, La, Cu, Cr, Ni and Zn in two sediment samples collected from the Ortigas river have been analysed using the three step sequential extraction procedure (SEP) described by BCR, Community Bureau of Reference (now Standards Measurements and Testing Programme) of the European Union. In order to perform a mass balance, a fourth step has been included, i.e. digestion of the residue from the third extraction step. ICP-AES was used for the determination of Cu, Cr, Ni, Zn, Th, Ce and La. Because of the inadequate sensitivity of ICP-AES for the analysis of uranium, ICP-MS was used for the determination of this element. Standard addition method was required, because of the matrix effects. Finally, SEP was applied to the Ortigas river sediments, showing that most of the elements were found mainly associated with the residual sediment fraction, except for uranium, which was found as an exchangeable species in approximately 60% of its total content. Recoveries of 86-108% have been obtained.
Related Concept Videos
Extraction: Advanced Methods
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)