A new type of radiofrequency capacitively coupled plasma torch is presented. The torch electrode geometry is coaxial with a tubular central electrode and one or two outer ring electrodes. The argon plasma is generated at 275 W radiofrequency power and 27.12 MHz and it has a very good stability and a low gas consumption of 0.4 l min(-1). The nebulized sample is introduced through the tubular electrode into the core of the annular shaped plasma thus achieving a better atomisation and a lower background. The limits of detection for 20 elements are in the range of ng ml(-1) and the dynamic range between 2.5 and 3.5. The best results are obtained with the torch with two outer ring electrodes.