Related Experiment Video
Updated: Jun 28, 2026

Purification of Human S100A12 and Its Ion-induced Oligomers for Immune Cell Stimulation
Published on: September 29, 2019
Flow injection dialysis for the determination of anions using ion chromatography
K Grudpan1, J Jakmunee, P Sooksamiti
1Department of Chemistry, Faculty of Science, Chiang Mai University, Chiang Mai 50200, Thailand.
Abstract:
Flow injection dialysis (FID) coupling with ion chromatography (IC) is proposed for simultaneous determination of some anions (bromide, chloride, fluoride, nitrate, nitrite, phosphate and sulfate). A standard or sample containing the anions is injected into a donor stream of a mixture (0.022 M Na(2)CO(3) and 0.028 M NaHCO(3)) flowing into a dialysis cell. The bolus of the dialysate containing the anions, in the acceptor stream of water, flows to the IC injection valve where a portion of the bolus is injected into the IC and analysed under normal IC conditions, with a conductivity detector. FID provides on-line separation and dilution of the analytes from matrix especially from some species such as proteins, surfactant, particulates which may cause damage to the IC columns. Prolongation of life-time of the IC columns is an additional advantage to others which will be discussed. On-line dialysis-IC was also investigated.
Related Concept Videos
Ion-Exchange Chromatography
Dialysis
Ion Exchange
EDTA: Indirect and Alkalimetric Titration
Analyte Adsorption and Distribution
Affinity Chromatography