Related Experiment Video
Updated: Jun 28, 2026

Asymmetrical Flow Field-Flow Fractionation for Sizing of Gold Nanoparticles in Suspension
Published on: September 11, 2020
Integrated flow injection-solid phase spectrophotometric determination of minoxidil
A Ruiz-Medina1, M L Fernández-de Córdova, A Molina-Díaz
1Department of Physical and Analytical Chemistry, Faculty of Experimental Sciences, University of Jaén, 23071 Jaén, Spain.
Abstract:
A flow-through solid phase spectrophotometric (SPS) sensing device is proposed for the determination of minoxidil. The analyte is concentrated on Sephadex SP-C25 ion-exchanger packed in a flow cell and it is monitored by UV-Vis spectrophotometry at 282 nm, without derivatization reaction. When a HCl (10(-2) mol l(-1))/NaCl (5x10 (-2) mol l (-1)) solution is used as carrier/desorbing solution, the sensor responds linearly in the measuring range of 0.2-7, 0.1-4 and 0.05-2 mug ml(-1) with detection limits of 60, 33 and 6 ng ml(-1) for 600, 1000 and 2000 mul of sample, respectively. The relative standard deviations (%) for these volumes are 0.38, 1.06 and 2.63, respectively. The method was satisfactorily applied to the determination of minoxidil in pharmaceutical preparations and the results were compared with those obtained by high performance liquid chromatography (HPLC).