Sol-gel-derived amperometric biosensor for hydrogen peroxide based on methylene green incorporated in Nafion film
1Laboratory of Electroanalytical Chemistry and National Analysis and Research Center of Electrochemistry and Spectroscopy, Changchun Institute of Applied Chemistry, Chinese Academy of Sciences, 159 Renmin St, Changchun 130022, PR China.
Abstract:
A hydrogen peroxide biosensor was fabricated by coating a sol-gel-peroxidase layer onto a Nafion-methylene green modified electrode. Immobilization of methylene green (MG) was attributed to the electrostatic force between MG(+) and the negatively charged sulfonic acid groups in Nafion polymer, whereas immobilization of horseradish peroxidase was attributed to the encapsulation function of the silica sol-gel network. Cyclic voltammetry and chronoamperometry were employed to demonstrate the feasibility of electron transfer between sol-gel-immobilized peroxidase and a glassy carbon electrode. Performance of the sensor was evaluated with respect to response time, sensitivity as well as operational stability. The enzyme electrode has a sensitivity of 13.5 muA mM(-1) with a detection limit of 1.0x10(-7) M H(2)O(2), and the sensor achieved 95% of the steady-state current within 20 s.