Related Experiment Video
Updated: Jun 28, 2026

Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
Flow injection hydride generation electrothermal atomic absorption spectrometry with in-atomizer trapping for the
J F Tyson1, R I Ellis, G Carnrick
1Department of Chemistry, University of Massachusetts, Box 34510, Amherst, MA 01003-4510, USA.
Abstract:
Lead hydride was generated from acid solution, containing potassium ferricyanide as an oxidizing agent, by the reaction with alkaline borohydride solution. The effects of reaction conditions (hydrochloric acid, ferricyanide and borohydride concentrations), and the lengths of reaction and stripping coils were studied. The effects of trapping temperature and argon flow rate were also investigated. Under the conditions giving the best peak area sensitivity, the detection limit (concentration giving a signal equal to three S.D. of the blank signal) was 0.12 mug l(-1) for a 1000 mul injection volume. The detection limit was improved to 0.03 mug l(-1) when the ferricyanide was purified by passage through a cation-exchange resin. Two calcium supplement materials were analyzed by the flow injection (FI)-hydride generation (HG)-electrothermal atomization atomic absorption spectrometry (ETAAS) method, giving values of 0.55 and 0.66 mug g(-1), in agreement with results obtained by previously validated methods. For a 500-mg sample the limits of detection and quantification were 0.006 and 0.02 mug g(-1), respectively.
More Related Videos
Related Concept Videos
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...