The effect of analyte concentration range on measurement errors obtained by NIR spectroscopy

M R Riley1, H M Crider

  • 1Room 403, Shantz Building, Department of Agricultural and Biosystems Engineering, University of Arizona, Tucson, AZ 85721, USA.

Talanta
|October 31, 2008
PubMed
Summary

Near infrared spectroscopy (NIRS) can quantify key cell culture analytes, but accuracy decreases at low concentrations (0-1 mM). Wider ranges (0-30 mM) offer better precision for noninvasive monitoring.

Related Concept Videos

NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences01:20

Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences

Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and refractory oxide ion...
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
Atomic Absorption Spectroscopy: Interference01:25

Atomic Absorption Spectroscopy: Interference

Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Drug Concentrations: Measurements01:23

Drug Concentrations: Measurements

Drug concentration is the quantity of a drug present in a biological sample. Measuring drug amounts in biological samples allows the clinician to understand how a drug is absorbed, distributed, metabolized, and excreted. Samples can be obtained through invasive or non-invasive methods. Invasive techniques involve surgical or parenteral interventions to gather blood, cerebrospinal fluid, or tissue biopsy. Conversely, non-invasive approaches provide samples like urine, feces, and saliva.
Plasma —...
Atomic Absorption Spectroscopy: Lab01:21

Atomic Absorption Spectroscopy: Lab

For AAS measurements, samples must be introduced as clear solutions, often requiring extensive preliminary treatment to dissolve materials like soils, animal tissues, and minerals. Common methods for sample preparation include treatment with hot mineral acids, wet ashing, combustion in closed containers, high-temperature ashing, or fusion with reagents.
 Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...