Urea biosensor based on amperometric pH-sensing with hematein as a pH-sensitive redox mediator
A Pizzariello1, M Stredanský, S Stredanská
1POLYtech, Area Science Park, Padriciano 99, 34122 Trieste, Italy.
Abstract:
The natural dye hematein in water solution was used as a pH-sensitive redox-active mediator for amperometric pH-sensing. The electrochemical characteristics were studied using cyclic voltammetry and chronoamperometry. Several types of urea biosensors were constructed with urease on the surface of platinum and graphite composite electrodes or in the bulk of the graphite composite. They were used for the amperometric urea determination at a working potential of 0 mV (versus SCE) using 0.5 mM hematein. Detection limits and response linearity was in the micromolar range depending on the biosensor type, concentration and pH of buffers used. An interference study of various cations, anions, and substances, which may be present in real samples demonstrated good selectivity for the determination of urea. The biosensors showed good operational (>3 h) and storage (>3 months) stability. The results of urea determination in blood and urine obtained by biosensor correlated well with those obtained by a spectrophotometric reference method.
Related Concept Videos
Amperometry: Overview
Titration of a Weak Base with a Strong Acid
Using the ICE table and substituting the Kb value, we calculate the initial pH of 50 mL of 0.1 M ammonia to be 11.11. Addition of 25 mL of 0.1 M hydrochloric acid to this solution of ammonia results in a buffer with an equal concentration of ammonia and ammonium ions. The pH of this buffer can be calculated by substituting these...