Related Experiment Video
Updated: Jun 28, 2026

Fluorescent Lateral Flow Immunoassay Based on Quantum Dots Nanobeads
Published on: June 28, 2024
High-performance liquid chromatographic assay of fleroxacin in human serum using fluorescence detection
P Djurdjevic1, M Jelikic-Stankov, A Laban
1Faculty of Science, P.O. Box 60, 34000 Kragujevac, Yugoslavia.
Abstract:
A HPLC method has been developed for the direct assay of fleroxacin in serum, without previous extraction. Serum samples, after the addition of sodium dodecylsulfate (0.5%), were injected directly into an LC Hisep column. The mobile phase consisted of acetonitrile, water and triethylamine in a per cent volume ratio 18:80:2. The pH of the mobile phase was adjusted to 6.50 with the addition of phosphoric acid. The drug was detected fluorometrically at lambda (ex )=280 nm and lambda (em )=450 nm . The linear concentration range of fleroxacin was between 0.01 and 2.0mg/l with a detection limit of 1ng/ml.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Immunofluorescence Microscopy
The...