Related Experiment Video
Updated: Jun 28, 2026

Chemiluminescence-based Assays for Detection of Nitric Oxide and its Derivatives from Autoxidation and Nitrosated Compounds
Published on: February 16, 2022
Flow injection analysis of trace amounts of NADH with inhibited chemiluminescent detection
1Department of Chemistry, Fuzhou University, Fuzhou, Fujian 350002, PR China.
Abstract:
The flow injection analysis of NADH was developed based on its inhibitory effect on a luminol-potassium hexacyanoferrate(III) chemiluminescence (CL) system. Under the optimum conditions, the decreased CL intensity was linear to the NADH concentration in the range of 5.0x10(-8)-2.0x10(-5) mol l(-1). The relative standard deviation was 1.24% for eight determinations of 1.0x10(-6) mol l(-1) NADH and the detection limit (3s(b1)/S) was 2.0x10(-8) mol l(-1). The study of interfering substances was also performed and the method was applied to the direct determination of trace amounts of NADH in synthetic samples with satisfactory results. Moreover, the inhibitory mechanism of CL was investigated with CL spectra, fluorescence spectra and UV-visible absorption spectra.
More Related Videos
08:25Measuring Nitrite and Nitrate, Metabolites in the Nitric Oxide Pathway, in Biological Materials using the Chemiluminescence Method
Published on: December 25, 2016
10:28A Semi-High-Throughput Adaptation of the NADH-Coupled ATPase Assay for Screening Small Molecule Inhibitors
Published on: August 17, 2019