Related Experiment Video
Updated: Jun 28, 2026

Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
Creatinine sensitive biosensor based on ISFETs and creatinine deiminase immobilised in BSA membrane
Alexey P Soldatkin1, Jean Montoriol, William Sant
1IFoS-UMR/CNRS 5621, Ecole Centrale de Lyon, 69131 36 Av. Guy de Collongue, F69134 Ecully, Cedex, France; Institute of Molecular Biology and Genetics, National Academy of Sciences of Ukraine, 150 Zabolotnogo St., Kiev 03143, Ukraine.
Abstract:
A creatinine sensitive biosensor based on ion sensitive field-effect transistors (ISFETs) with immobilised creatinine deiminase (CD) is developed. CD is immobilised on the transducer surface by classical cross-linking with bovine serum albumin (BSA) in a glutaraldehyde (GA) vapour. The linear dynamic ranges of biosensors are between 0 and 5 mM creatinine concentration, and the sensor sensitivity depends on the sample buffer concentration. Minimal detection limit for creatinine determination in model solution with 144 mM NaCl and 5% BSA, pH 7.4, is about 10 muM. Biosensor responses are reproducible and stable during continuous work at least for 8 h, and the relative standard deviation of sensor response is approximately 3% (n=48, for creatinine concentrations of 0.2 and 0.6 mM). An investigation about storage stability of creatinine sensitive ENFETs kept in dry at 4-6 degrees C shows that biosensors demonstrate an excellent storage stability for at least 6 months and more. Moreover, creatinine sensitive enzymatic field-effect transistors (ENFETs), demonstrating very good performances, are very selective and specific and well suitable for hemodialysis monitoring.
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Microbial Biosensors