Fluorimetric determination of fluoride in a flow assembly integrated on-line to an open/closed FIA system to remove
M Garrido1, A G Lista, M Palomeque
1Department of Chemistry, FIA Laboratory, Universidad Nacional del Sur, Av. Alem 1253, 8000, Bahia Blanca, Argentina.
Abstract:
A simple flow injection fluorimetric method for fluoride determination is proposed. The method is based on the enhanced fluorescence of quercitin-Zr(IV) complex when fluoride ion is present in the sample. An open/closed FIA manifold with a mini-column of Dowex 50W X8 resin was used to remove the most important interference (aluminum). The two FIA assemblies were integrated on-line to automate the pretreatment of the water sample and fluoride determination. The calibration graph was linear over the range 0.1-3.0 mug ml(-1) of fluoride with a correlation coefficient of 0.999 and LOD 0.06 mug ml(-1). The relative standard deviation was 2.5% and the sample throughput was 52 h(-1) without pretreatment and 10 h(-1) with pretreatment of the sample. The method was applied to the determination of fluoride in water samples.
More Related Videos
09:04Identifying Per- and Polyfluorinated Chemical Species with a Combined Targeted and Non-Targeted-Screening High-Resolution Mass Spectrometry Workflow
Published on: April 18, 2019
09:51TD-DFT Guided Advanced E-Eye Sensing Technique for On-site Quantification of Fe, Cr, F, and As in the Environmental, Biological, and Food Samples
Published on: September 19, 2025
Related Concept Videos
Atomic Fluorescence Spectroscopy
Gas Chromatography: Types of Detectors-II