Gliclazide as novel carrier in construction of PVC-based La(III)-selective membrane sensor
Mohammad Reza Ganjali1, Azadeh Daftari, Morteza Rezapour
1Department of Chemistry, Tehran University, P.O. Box 14155 6455, Tehran, Iran.
Abstract:
A PVC-based sensor for La(3+) ions based on N-[hexahydrocyclopentapyrol-2((1H)yl)amino]carbonyl]-4-methyl benzene sulfonamide (gliclazide) as a novel carrier was prepared. The electrode exhibits a Nernstian response for La(3+) over a wide concentration range (1.0x10(-1)-1.0x10(-6) M) with a slope of 20.1 mV per decade. The limit of detection is 8.0x10(-7) M. The sensor has a very short response time (<15 s) and a useful working pH range of 4.0-8.0. The proposed membrane sensor shows excellent discriminating ability towards La(3+) ions with regard to several alkali, alkaline earth, transition and heavy metal ions. The electrode was successfully applied for determination of La(3+) in binary mixtures.