Related Experiment Video
Updated: Jun 28, 2026

Quantification of Metal Leaching in Immobilized Metal Affinity Chromatography
Published on: January 17, 2020
Detection of metal ions using ion-channel sensor based on self-assembled monolayer of thioctic acid
Ruphino Zugle1, James Kambo-Dorsa, Victor Patrick Yao Gadzekpo
1Department of chemistry, University of Cape Coast, Cape Coast, Ghana.
Abstract:
Gold electrodes were chemically modified with thioctic acid monolayer designed to mimic biological ion-channel membranes. The technique was then used in the determination of alkali, alkaline earth, thallium(I), and lanthanum metal cations as analytes. Cyclic voltammograms (CV) of [Fe(CN)6](3-) an electroactive marker, were measured in the presence of the various types of analyte cations. In the absence of the analyte cation, electrostatic repulsion between the marker anions and the carboxylate groups of the receptor monolayer hindered the approach of the marker anion to the electrode surface and hence hindered its reduction. The modified electrodes responded well to the metal cations except the alkali metal cations. The sensors could detect the trivalent cation La(3+) at concentrations as low as 10(-8) M. The response of the sensor to the metal cations increase in the order alkali metal
More Related Videos
11:04Ion Mobility-Mass Spectrometry Techniques for Determining the Structure and Mechanisms of Metal Ion Recognition and Redox Activity of Metal Binding Oligopeptides
Published on: September 7, 2019
10:31Detection and Recovery of Palladium, Gold and Cobalt Metals from the Urban Mine Using Novel Sensors/Adsorbents Designated with Nanoscale Wagon-wheel-shaped Pores
Published on: December 6, 2015