Related Experiment Video
Updated: Jun 28, 2026

Single Liposome Measurements for the Study of Proton-Pumping Membrane Enzymes Using Electrochemistry and Fluorescent Microscopy
Published on: February 21, 2019
Determination of l-carnitine using enantioselective, potentiometric membrane electrodes based on macrocyclic
Alexander A Rat'ko1, Raluca-Ioana Stefan, Jacobus Koos F van Staden
1Department of Chemistry, University of Pretoria, Pretoria 0002, South Africa.
Abstract:
In order to determine the enantiopurity of l-carnitine three enantioselective, potentiometric membrane electrodes were proposed for the assay of l-carnitine. The electrodes were designed using macrocyclic glycopeptide antibiotics-vancomycin and teicoplanin. Acetonitrile was added to the teicoplanine to design a modified teicoplanine based electrode. The linear concentration ranges for the proposed enantioselective membrane electrodes were 10(-4) to 10(-2)moll(-1) for electrodes based on vancomycin and teicoplanin and 10(-5) to 10(-2)moll(-1) for electrode based on teicoplanin modified with acetonitrile. The slopes of the electrodes were 56.5mV per pl-carnitine; 54.5mV per pl-carnitine and 58.3mV per pl-carnitine for vancomycin-, teicoplanin- and teicoplanin modified with acetonitrile-based electrodes, respectively. The enantioselectivity was determined over d-carnitine. The proposed electrodes could be employed reliably for the assay of l-carnitine raw material and its pharmaceutical formulation, Carnilean((R)) capsules. The surfaces of the electrodes are stable and easily renewable by polishing on alumina paper.