Manganese detection in marine sediments: anodic vs. cathodic stripping voltammetry
Related Experiment Video
Updated: Jun 28, 2026

Clean Sampling and Analysis of River and Estuarine Waters for Trace Metal Studies
Published on: July 1, 2016
Craig E Banks1, Jaanus Kruusma, Ryan R Moore
1Physical and Theoretical Chemistry Laboratory, Oxford University, South Parks Road, Oxford OX1 3QZ, UK.
Abstract:
Three different electroanalytical techniques for the detection of manganese in marine sediments are evaluated. The anodic stripping voltammetry of manganese at an in situ bismuth-film-modified boron-doped diamond electrode and cathodic stripping voltammetry at a carbon paste electrode are shown to lack the required sensitivity and reproducibility whereas cathodic stripping voltammetry at a bare boron-doped diamond electrode is shown to be reliable and selective with a limit of detection, from applying a 60s accumulation period of 7.4 x 10(-7)M and a sensitivity of 0.24AM(-1). The method was used to evaluate the manganese content of marine sediments taken from Sibenik, Croatia.
More Related Videos
Related Concept Videos
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Voltammetric Techniques: Linear-Scan (E vs Time)
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
Electrodeposition
Electrodeposition can...
Voltammetric Techniques: Cyclic Voltammetry
Voltammetric Techniques: Pulse Voltammetry