Determination of terbinafine in pharmaceuticals and dialyzates by capillary electrophoresis
Peter Mikus1, Iva Valásková, Emil Havránek
1Department of Pharmaceutical Analysis and Nuclear Pharmacy, Faculty of Pharmacy, Comenius University, Odbojárov 10, SK-832 32 Bratislava, Slovak Republic.
Abstract:
A capillary electrophoresis method has been developed for the separation and determination of terbinafine (TER) in various pharmaceutically relevant matrices. Capillary zone electrophoresis (CZE) separation and UV absorbance photometric detection were carried out in a 160mm capillary tube with a 300mum i.d., hydrodynamically (membrane) closed. The influences of pH, carrier cation and counterion on migration parameters of TER were studied and the following conditions were selected: a 20mmoll(-1) glycine running buffer adjusted to pH 2.7 with acetic acid, 0.2% (w/v) methylhydroxyethylcellulose (m-HEC) as an electro-osmotic flow (EOF) suppressor, a 250muA driving current, and 20 degrees C. The optimized separation conditions were convenient for the determination of TER in commercial tablets and spray and in dialyzates. Here, the dialysis was used to investigate in vitro permeation of TER through the skin from the gel. The samples of dialyzates were examined with and without simple extraction procedure and the results were compared. A permeation profile of the drug present in the gel of given composition was obtained analyzing pretreated samples. The proposed electrophoretic method was successfully validated. It was suitable for the simple, sensitive, rapid and highly reproducible assay of TER. CZE analysis was completed within 5.5min. The detection limit of TER was 1.73mumoll(-1) at a 224nm detection wavelength. The intra- and inter-laboratory precisions over the concentration range 6.0-60.0mumoll(-1) were between 0.32-0.69% and 1.04-1.44% including R.S.D. of migration times and peak areas, respectively. The mean absolute recoveries of drugs from samples were found to be 98.34 (tablets) and 99.47% (spray). It is suggested that there are potentialities to determine TER present in unpretreated complex samples, as CZE in a hydrodynamically closed separation system may be easily on-line combinable with purification and preconcentration CE modes (e.g., isotachophoresis, ITP).
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Electrophoresis: Overview
There...
Capillary Electrophoresis: Instrumentation