Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
Determination of arsenic compounds in beverages by high-performance liquid chromatography-inductively coupled plasma
N M M Coelho1, L M Coelho, E S de Lima
1Institute of Chemistry, University of Uberlândia, Av. João Naves de Avila, 2160, Campus Santa Mônica, CEP: 38.408.110 Uberlândia, MG, Brazil.
Abstract:
Arsenic compounds including arsenous acid (As(III)), arsenic acid (As(V)), dimethylarsinic acid (DMA) and monomethylarsonic acid (MMA) were separated by high-performance liquid chromatography (HPLC) and detected by inductively coupled plasma mass spectrometry (ICP-MS). A Hamilton PRX-100 anionic-exchange column and a pH 8.5 K(2)HPO(4)/KH(2)PO(4) 5.0x10(-3)molL(-1) mobile phase were used to achieve arsenic speciation. The separation of arsenic species provided peaks of As(III) at 2.75min, DMA at 3.33min, MMA at 5.17min and As(V) at 12.5min. The detection limits, defined as three times the standard deviation of the lowest standard measurements, were found to be 0.2, 0.2, 0.3 and 0.5ngmL(-1) for As(III), DMA, MMA and As(V), respectively. The relative standard deviation values for a solution containing 5.0mugL(-1) of As(III), DMA, MMA and As(V) were 1.2, 2.1, 2.5 and 3.0%, respectively. This analytical procedure was applied to the speciation of arsenic compounds in drinking (soft drink, beer, juice) samples. The validation of the procedure was achieved through the analysis of arsenic compounds in water and sediment certified reference materials.
More Related Videos
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
High-Performance Liquid Chromatography: Types of Detectors
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...