Alizarin violet modified carbon paste electrode for the determination of trace silver(I) by adsorptive voltammetry
Yi-Heng Li1, Hong-Qi Xie, Fang-Qin Zhou
1College of Chemistry, Xiangtan University, Xiangtan 411105, Hunan, P.R. China.
Abstract:
A sensitive and selective procedure for the determination of trace silver(I) at alizarin violet modified carbon paste electrode has been developed. Silver(I) was accumulated on the electrode surface via complex formation with modifier in 0.1mol/L acetate buffer (pH 5.2). After electrochemical reduction of silver(I) had been carried out, the reoxidation wave of silver(0) appeared at 0.05V (versus SCE) on scanning the potential in the positive direction in 0.1mol/L H(2)SO(4)+2.0x10(-4)mol/L KBr. For a preconcentration time of 3min, the detection limit is 1.0x10(-10)mol/L and the linear range is from 3.0x10(-10) to 1.2x10(-7)mol/L. Many coexisting metal ions have little or no effect on the determination of silver(I). The proposed method was applied to the determination of silver in waste water and zinc alloy samples with satisfactory results.
Related Concept Videos
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Precipitation Titration Curve: Analysis
Electrodeposition
Electrodeposition can...
Electrogravimetric Analysis: Overview
To test the completeness of the...
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...