Cobalt as internal standard for arsenic and selenium determination in urine by simultaneous atomic absorption
Paulo R M Correia1, Pedro V Oliveira
1Escola de Artes, Ciências e Humanidades, Universidade de São Paulo, Av Arlindo Bettio 1000, CEP 03828-000, São Paulo, SP, Brazil.
Abstract:
The effectiveness of internal standardization for simultaneous atomic absorption spectrometry (SIMAAS) was investigated for As and Se determination in urine. Co and Sn were selected as internal standard (IS) candidates based on the evaluation of some physico-chemical parameters related to the atomization. Correlation graphs, plotted from the normalized absorbance signals (n=20) of internal standard (axis y) versus analyte (axis x), precision, and accuracy of the analytical results were the supportive parameters to choose Co as the most appropriate IS. The urine samples were diluted 1+2 to 1.0% (v/v) HNO(3)+80mugL(-1) Co(2+). The mixture 20mug Pd+3mug Mg was used as chemical modifier and the optimized temperatures for pyrolysis and atomization steps were 1400 and 2300 degrees C, respectively. The characteristic masses for As (47+/-1pg) and Se (72+/-2pg) were estimated from the analytical curves. The detection limits (n=20, 3delta) were 1.8+/-0.1 and 2.6+/-0.1mugL(-1) for As and Se, respectively. The reliability of the entire procedure was checked with the analysis of certified reference material from Sero AS(Seronormtrade mark Trace Elements in Urine). The obtained results showed the matrix interference disallowed the instrument calibration with aqueous standards. The best analytical condition was achieved when matrix-matched standards were used in combination with Co as IS, which improved the recoveries obtained for As. Under this experimental condition, eight urine samples were analysed and spiked with 10 and 25mugL(-1) As and Se. The mean recoveries were 96+/-6% (10mugL(-1) As), 95+/-6% (25mugL(-1) As), 101+/-7% (10mugL(-1) Se), and 97+/-4% (25mugL(-1) Se).
More Related Videos
08:21Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
06:00Biological Samples Preparation for Speciation at Cryogenic Temperature using High-Resolution X-Ray Absorption Spectroscopy
Published on: May 27, 2022
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Precipitation and Co-precipitation
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...