The effects of sampling materials selection in the collection of reduced sulfur compounds in air
Ki-Hyun Kim1, Gyoo-Hoon Choi, Ye-Jin Choi
1Department of Earth and Environmental Sciences, Sejong University, Seoul, Republic of Korea.
Abstract:
In this study, the analytical bias in the measurements of reduced sulfur compounds (RSC) was investigated in terms of sorptive loss caused by the materials selected for the sample introduction. For the purpose of this study, three vacuum samplers made in the combination of different vacuuming efficiencies (e.g., rapid versus slow sampling) and different materials (i.e., Teflon versus stainless steel (SS)) were tested to evaluate the sampling recovery rate (RR) for five RSCs: H(2)S, CH(3)SH, DMS, CS(2) and DMDS. To make a parallel comparison of RR, the RSC standard samples contained in one bag were transferred to another bag using each sampling system. Their relative contents between, before, and after the transfer were then evaluated between different samplers to assess the sampling bias caused by the interaction between RSC and the sampling material. In the case of the most reactive compound, H(2)S, the sampling loss from the SS inlet line amounted to as high as 45%, while that for the Teflon counterpart was almost insignificant. When the sampling time was arbitrarily elongated (i.e., use of a slow sampler), the sampling loss rate of the SS inlet sampler became more significant with the RR values dropping down from 55 to 70%, across different RSCs. The overall results of our comparative study indicate that the sampling system for the reactive gaseous compounds should be checked for the material feasibility to guarantee sufficient analytical reliability.
More Related Videos
09:31Preparation of Authigenic Pyrite from Methane-bearing Sediments for In Situ Sulfur Isotope Analysis Using SIMS
Published on: August 31, 2017
08:37Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
Published on: December 10, 2015
Related Concept Videos
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Sulfur Assimilation
Microbes and the Sulfur Cycle
Sampling Methods: Overview
In analytical chemistry, the choice of sampling...