Related Experiment Video
Updated: Jun 28, 2026

Quantification of three DNA Lesions by Mass Spectrometry and Assessment of Their Levels in Tissues of Mice Exposed to Ambient Fine Particulate Matter
Published on: May 29, 2019
Method optimization for quantitative analysis of octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine (HMX) by liquid
Xiaoping Pan1, Kang Tian, Lindsey E Jones
1The Institute of Environmental and Human Health (TIEHH), Department of Environmental Toxicology, Texas Tech University, Lubbock, TX 79409-1163, USA.
Abstract:
A simple, sensitive LC-ESI-MS method was optimized for quantitative analysis of octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine (HMX) in environmental samples. Under negative ionization mode, HMX can form adduct ions with various organic acids and salts, including acetic acid, formic acid, propionic acid, ammonium nitrate, ammonium chloride, sodium nitrite, and sodium nitrate. Acetic acid was chosen as additive and the ion, [M+CH(3)COO](-) with m/z=355 was used for selective ion monitoring (SIM) in this study. Good sensitivity was achieved with low acetic acid concentration in the mobile phase and relatively low capillary temperature. The method detection limit was 0.78pg for HMX in standard solution. Linearity (R(2)>0.9998) was obtained at low concentrations (0.5-50mug/L). This method has been used to determine HMX concentrations in water samples and lizard egg samples from an animal exposure study.