Experimental contamination of a closed endotracheal suction system: 24 h vs 72 h
E Meyer1, M Schuhmacher, W Ebner
1Department of Environmental Health Sciences, University Medical Center Freiburg, Freiburg, Germany.
Objective:
To ascertain the desirability of replacing closed suction systems after 72 h rather than after 24 h (manufacturer's recommendations) because it is possible that a reduction in the frequency of manipulations might reduce the risk of exogenous nosocomial pneumonia. We investigated the presence of time-dependent differences (after 24 h and 72 h) in pathogen survival/growth in artificially contaminated closed suction catheters (OptiFlo).
Design:
The trial simulated bacterial contamination of the airways using a suspension of 2 x 10(3) CFU/ml of Staphylococcus aureus or Pseudomonas aeruginosa. Contamination was performed on a total of 80 catheters. Forty were contaminated a total of eight times every 45-60 min. Another 40 catheters underwent the same procedure 24 times over three consecutive days. Microbiological analysis of the catheters took place after 24 h and 72 h, respectively.
Results:
The mean S. aureus load was 9.4 CFU/catheter after eight suction procedures and 6.2 CFU/catheter after 24 suction procedures (3 days). Mean growth of P. aeruginosa was 5.3 CFU/catheter, and 8.2 CFU/catheter after 3 days. There was no statistically significant difference between day 1 and 3 for S. aureus (p = 0.474), but there was for P. aeruginosa (p = 0.004).
Conclusion:
Our findings show that, from an experimental point of view, it remains controversial whether routine change of closed suction catheters can be extended from 24 h to 72 h. However, clinical evidence suggests that prolonged use of a closed suctioning system is safe.
Related Concept Videos
Tracheostomy Suctioning II: Procedure
Sputum Studies II: Culture and Sensitivity
Sputum culture and sensitivity is a medical procedure used to diagnose bacterial infections in the respiratory tract and select the most appropriate antibiotics for treatment. This process involves analyzing sputum samples of thick and opaque secretions produced in the lungs and airways. These samples are collected from patients and then sent to the laboratory for analysis.
The test can identify various pathogens responsible for respiratory infections, including Streptococcus,...
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...