Needle position during (125)I seed implantation: accurately recognized by sagittal transrectal ultrasonography
Hiromichi Ishiyama1, Shouko Kotani, Takefumi Satoh
1Department of Radiology, Kitasato University School of Medicine, 1-15-1 Kitasato, Sagamihara, 228-8555, Japan. hishiyam@kitasato-u.ac.jp
Purpose:
The aim of this study was to assess the variation of probe rotation angles for detecting a single needle using sagittal images of transrectal ultrasonography (TRUS).
Materials And Methods:
A phantom study was performed. One needle was inserted through each of 10 holes of the template, and variations in the probe rotation angles for detecting the needle were measured.
Results:
The mean variation of probe rotation for detecting a single needle was 17.0 degrees (range 4 degrees -25 degrees ). Slightly broader variation was seen for the needle in holes farther away from the probe.
Conclusion:
Probe rotation angles for detecting a single needle displayed considerable variation. Seed locations recognized on sagittal imaging by TRUS are thus indeterminate, and real-time dose calculations using TRUS for (125)I seed implantation should be used with care.