An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Selection of the implant transgingival height for optimal biomechanical properties: a three-dimensional finite
Yingying Sun1, Liang Kong, Kaijin Hu
1Department of Oral and Maxillofacial Surgery, School of Stomatology, the Fourth Military Medical University, Xi'an, Shanxi, PR China.
Abstract:
We evaluated the effects of the transgingival height of an implant on the maximum equivalent stress in jaw bones and the maximum displacement in implant-abutment complex by a finite element method. The transgingival height ranged from 1.0-4.0 mm. Under axial load, the maximum equivalent stress in the cortical bone could be reduced by up to 4.7%, and under a buccolingual load, the maximum equivalent stresses in the cortical and the cancellous bones could be reduced by 17.3% and 18.5%, respectively. The maximum displacement of the implant-abutment complex could be reduced by 4.1% and 48.9% under axial and buccolingual loads, respectively. When the transgingival height was in the range of 1.7-2.8 mm, there was minimum stress in the jaw bones and minimum displacement in the implant-abutment complex. Data indicated that transgingival height played a more important part in protecting a dental implant under a buccolingual load than under an axial load; and transgingival heights ranging from 1.7-2.8 mm were biomechanically optimal for a screwed implant.