Related Experiment Video
Updated: Jun 28, 2026

Creating Rapid Oxygen Oscillations in Microbial Single-cell Growth Analysis using a Microfluidic Double-layer Device
Published on: July 18, 2025
Concentration determination of oxygen nanobubbles in electrolyzed water
Kenji Kikuchi1, Aoi Ioka, Takeo Oku
1Department of Materials Science, University of Shiga Prefecture, Hassaka, Hikone, Shiga 522-8533, Japan. kikuchik@mat.usp.ac.jp
Abstract:
Water electrolysis is well known to produce solutions supersaturated with oxygen. The oxygen in electrolyzed solutions was analyzed with a dissolved oxygen meter and the Winkler method of chemical analysis. The concentration of oxygen measured with the dissolved oxygen meter agreed with that obtained using the Winkler method. However, measurements using a 10-fold dilution method showed a larger concentration of dissolved oxygen compared to the above methods. We developed a modified Winkler method to measure total oxygen concentration more accurately, which agreed with the results obtained from the 10-fold dilution experiment. The difference in measurements is due to the existence of oxygen nanobubbles, as confirmed by the observation of dynamic light scattering using a laser. Further analysis of the oxygen nanobubbles demonstrated that the stability of the nanobubbles was sufficient for chemical reaction and solvation to bulk solution.
More Related Videos
Related Concept Videos
Concentration Cells
Consider the following voltaic cell:
Concentration Cells
Calculating Equilibrium Concentrations
A more...