A fortuitous syncope. The pitfalls of integrated bipolar defibrillator leads
Tushar V Salukhe1, Ian Wright, Matthew Wright
1Department of Cardiac Electrophysiology, St Mary's Hospital and Imperial College, London, UK. salukhe@aol.com
Abstract:
Myopotential oversensing in implantable defibrillators causing inhibition of pacing and inappropriate therapies is well described. Current literature is dominated by reports of diaphragmatic muscle as the source of such far-field oversensing. Those reporting pectoral muscle sources were invariably due to unipolar sensing circuits, incorrect DF-1 connections or inappropriate programming. We report an interesting case of pectoral muscle myopotential oversensing causing inhibition of bradycardia pacing leading to presyncope and syncope.