Validation and comparison of two multiplex technologies, Luminex and Mesoscale Discovery, for human cytokine
Ferdousi Chowdhury1, Anthony Williams, Peter Johnson
1Cancer Sciences Division, CRUK Clinical Centre, Somers Cancer Research Building, University of Southampton, Southampton General Hospital, Tremona Road, Southampton, SO16 6YD, England, United Kingdom.
Abstract:
Biomarker research has rapidly expanded over recent years aided by the progressive development of research tools, in particular the different multiplex technologies allowing simultaneous measurement of multiple analytes. It is foreseeable that such technology will have an integral role in clinical studies for establishing biomarker profiles of disease status, but validation of the tools is essential to confirm the reliability of their application. More comparable studies between multiplex platforms are required to enable users to determine which of these are best for a particular clinical study, as different platforms will have varying levels of performance for the validation parameters. Comparison of two multiplex platforms, the Luminex and the Mesoscale Discovery, has been performed to determine their performance for the validation parameters of sensitivity, precision and accuracy for the cytokines IL-2, IL-4, IL-8, IL-10, IL-12, IFNgamma and TNFalpha. When measuring high concentrations both platforms show good accuracy (within +/-25% recovery) with all cytokines except IL-12 for the MSD. At low concentrations, +/-25% recovery was seen with all cytokines except IL-2 and IL-8 for the Luminex and IL-2 and IL-12 for the MSD. Although quantitative differences are found, relative differences are comparable, and consequently both platforms have been shown to be suitable for analyzing trends in multiple cytokine profiles, with the Luminex having better precision and the Mesoscale Discovery having greater sensitivity.
More Related Videos
11:00Multiplex Cytokine Profiling of Stimulated Mouse Splenocytes Using a Cytometric Bead-based Immunoassay Platform
Published on: November 9, 2017
06:37Using Reference Reagents to Confirm Robustness of Cytokine Release Assays for the Prediction of Monoclonal Antibody Safety
Published on: September 15, 2023