Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
Double Resonance Techniques: Overview01:12

Double Resonance Techniques: Overview

Double resonance techniques in Nuclear Magnetic Resonance (NMR) spectroscopy involve the simultaneous application of two different frequencies or radiofrequency pulses to manipulate and observe two distinct nuclear spins. One important application of double resonance is spin decoupling, which selectively suppresses coupling with one type of nucleus while observing the NMR signal from another nucleus, simplifying the spectrum and enhancing resolution.
Spin decoupling is usually achieved by...
Phase Contrast and Differential Interference Contrast Microscopy01:26

Phase Contrast and Differential Interference Contrast Microscopy

Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences01:17

NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences

A pulse is a short burst of radio waves distributed over a range of frequencies that simultaneously excites all the nuclei in the sample. Upon passing a radio frequency pulse along the x-axis, the nuclei absorb energy corresponding to their Larmor frequencies and achieve resonance. This shifts the net magnetization vector from the z-axis toward the transverse plane. This angle of rotation of the magnetization vector, or the flip angle, is proportional to the duration and intensity of the pulse.
¹H NMR Signal Multiplicity: Splitting Patterns01:13

¹H NMR Signal Multiplicity: Splitting Patterns

When protons A and X are coupled, their nuclear spin energy levels are slightly modified. This is because the energy required to excite proton A to a spin state parallel to proton X is slightly different from the energy required for it to become anti-parallel to spin X. Consequently, there are two possible excitation frequencies for A (A1 and A2), depending on the spin state of X, and vice versa. The mutual nature of coupling implies that the difference between frequencies A1 and A2, indicated...
Two-Dimensional (2D) NMR: Overview01:12

Two-Dimensional (2D) NMR: Overview

The 1D NMR spectrum of large and complex molecules like natural products has complicated splitting patterns and overlapping signals, which can be easily interpreted using 2-dimensional (2D) NMR. Unlike 1D NMR, 2D NMR has two frequency axes that provide the coupling information between the nucleus A and nucleus B in a molecule. The process from which 2D spectra are obtained has four steps.
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse.

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Modulation of Cilia Motility by Vortex-Ultrasound-Induced Shear Stress.

ACS nano·2026
Same author

Quantitative ultrasound scatteromics for characterization of reconstructed neopharyngeal tissue: Assessment of postoperative speech outcome after total laryngopharyngectomy.

Ultrasonics·2026
Same author

High-contrast ultrafast ultrasound power doppler imaging using subarray adaptive temporal autocorrelation with phase compensation.

Biomedical physics & engineering express·2026
Same author

Ultrasound synthetic aperture imaging with SpecularNet for enhanced needle visualization.

Journal of medical ultrasonics (2001)·2026
Same author

Optimization of Ultrasound Phase-Encoded Multiplane Wave Tissue Harmonic Imaging.

Ultrasonic imaging·2026
Same author

Beyond total release: Fragment-resolved metrics for ultrasound microbubble-mediated drug delivery.

Biomedicine & pharmacotherapy = Biomedecine & pharmacotherapie·2026

Related Experiment Video

Updated: Jun 28, 2026

Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples
07:01

Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples

Published on: June 9, 2016

Dual high-frequency difference excitation for contrast detection.

Chih-Kuang Yeh1, Shin-Yuan Su, Che-Chou Shen

  • 1Department of Biomedical Engineering and Environmental Sciences, National Tsing Hua University, Hsinchu, Taiwan, ROC.

IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
|November 7, 2008
PubMed
Summary

A novel dual-frequency ultrasound technique efficiently generates low-frequency nonlinear scattering from microbubbles. This method enhances microbubble detection and contrast-to-tissue ratios for improved diagnostic imaging.

More Related Videos

Local Field Fluorescence Microscopy: Imaging Cellular Signals in Intact Hearts
10:33

Local Field Fluorescence Microscopy: Imaging Cellular Signals in Intact Hearts

Published on: March 8, 2017

Excitation-Scanning Hyperspectral Imaging Microscopy to Efficiently Discriminate Fluorescence Signals
07:34

Excitation-Scanning Hyperspectral Imaging Microscopy to Efficiently Discriminate Fluorescence Signals

Published on: August 22, 2019

Related Experiment Videos

Last Updated: Jun 28, 2026

Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples
07:01

Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples

Published on: June 9, 2016

Local Field Fluorescence Microscopy: Imaging Cellular Signals in Intact Hearts
10:33

Local Field Fluorescence Microscopy: Imaging Cellular Signals in Intact Hearts

Published on: March 8, 2017

Excitation-Scanning Hyperspectral Imaging Microscopy to Efficiently Discriminate Fluorescence Signals
07:34

Excitation-Scanning Hyperspectral Imaging Microscopy to Efficiently Discriminate Fluorescence Signals

Published on: August 22, 2019

Area of Science:

  • Biomedical Engineering
  • Acoustics
  • Medical Imaging

Background:

  • High-frequency nonlinear oscillations of ultrasound contrast agents aid microbubble distinction.
  • Inefficient oscillations and harmonic attenuation limit current high-frequency ultrasound detection methods.

Purpose of the Study:

  • To develop a dual-frequency difference excitation technique for efficient low-frequency nonlinear scattering from microbubbles using high-frequency ultrasound.
  • To overcome limitations of high-frequency harmonic generation in ultrasound contrast imaging.

Main Methods:

  • Proposed a dual-frequency excitation pulse using two high-frequency sinusoids with a frequency difference matching microbubble resonance.
  • Utilized the envelope frequency of the excitation pulse to stimulate nonlinear microbubble oscillations.
  • Conducted hydrophone measurements and phantom experiments with speckle-generating flow phantoms.

Main Results:

  • The dual-frequency envelope effectively induced significant nonlinear scattering from microbubbles, particularly near their resonance frequency.
  • Induced nonlinear responses increased with pulse pressures.
  • Achieved contrast-to-tissue ratios of up to 26 dB for second-order and 36 dB for fourth-order nonlinear responses.

Conclusions:

  • The dual-frequency difference excitation technique efficiently generates low-frequency nonlinear scattering from microbubbles.
  • This method offers improved contrast-to-tissue ratios and retains high imaging resolution.
  • Potential applications include microbubble fragmentation and cavitation using high-frequency ultrasound.