Imaging performance with different doses in skeletal radiography: comparison of a needle-structured and a
Stefan Wirth1, Marcus Treitl, Maximilian F Reiser
1Department of Clinical Radiology, Ludwig-Maximilians-University, Nussbaumstr 20, 80336 Munich, Germany. Stefan.Wirth@med.uni-muenchen.de
Purpose:
To evaluate possible radiation dose reduction in the extremities with use of digital radiography and a needle-structured image plate (NIP) by comparing this technique with digital radiography performed with a powder-structured image plate (PIP) and a flat-panel detector (FPD).
Materials And Methods:
This study was approved by the local review board. A total of 72 plain radiographs of the feet of six human cadavers were obtained with four surface entrance doses (65, 43, 20, and 10 micro Gy) by using three systems. The reference image of each specimen was obtained with an 85-micro Gy dose and with use of a PIP. Five independent blinded radiologists evaluated the images. The noise level and the depiction of the cortical bone, trabecular bone, and soft tissue were rated and compared with those of the reference image by using a five-point scale. An overall image score was developed for these four criteria by calculating the unweighted mean. The Wilcoxon test was used to assess differences between overall image scores.
Results:
For each dose, NIP images were significantly superior (P < .001), whereas FPD images and PIP images were significantly inferior (P < .01). NIP images obtained with 65-, 43-, and 20-micro Gy doses were significantly superior to reference images and to FPD and PIP images obtained with a 65-micro Gy dose. There were no significant differences between reference images and FPD images obtained with 65- and 43-micro Gy doses.
Conclusion:
Radiation dose can be reduced by 75% in clinical skeletal imaging of peripheral extremities by using NIP, with no significant loss of information. For FPD images, this might be possible with a dose reduction of 50%.
More Related Videos
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
X-ray Imaging
Determination of Crystal Structures
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...