Assessment of dosimetrical performance in 11 Varian a-Si-500 electronic portal imaging devices
Awusi Kavuma1, Martin Glegg, Garry Currie
1Department of Radiotherapy Physics, Beatson West of Scotland Cancer Center, NHS Greater Glasgow & Clyde, 1053 Great Western Road, Glasgow G12 0YN Scotland, UK. kavuma.awusi@northglasgow.scot.nhs.uk
Abstract:
Dosimetrical characteristics of 11 Varian a-Si-500 electronic portal imaging devices (EPIDs) in clinical use for periods ranging between 10 and 86 months were investigated for consistency of performance and portal dosimetry implications. Properties studied include short-term reproducibility, signal linearity with monitor units, response to reference beam, signal uniformity across the detector panel, signal dependence on field size, dose-rate influence, memory effects and image profiles as a function of monitor units. The EPID measurements were also compared with those of the ionization chambers' to ensure stability of the linear accelerators. Depending on their clinical installation date, the EPIDs were interfaced with one of the two different acquisition control software packages, IAS2/IDU-II or IAS3/IDU-20. Both the EPID age and image acquisition system influenced the dosimetric characteristics with the newer version (IAS3 with IDU-20) giving better data reproducibility and linearity fit than the older version (IAS2 with IDU-II). The relative signal response (uniformity) after 50 MU was better than 95% of the central value and independent of detector. Sensitivity for all EPIDs reduced continuously with increasing dose rates for the newer image acquisition software. In the dose-rate range 100-600 MU min(-1), the maximum variation in sensitivity ranged between 1 and 1.8% for different EPIDs. For memory effects, the increase in the measured signal at the centre of the irradiated field for successive images was within 1.8% and 1.0% for the older and newer acquisition systems, respectively. Image profiles acquired at a lower MU in the radial plane (gun-target) had gradients in measured pixel values of up to 25% for the older system. Detectors with software/hardware versions IAS3/IDU-20 have a high degree of accuracy and are more suitable for routine quantitative IMRT dosimetrical verification.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...