The noninvasive estimation of central aortic blood pressure in patients with aortic stenosis
Ronak Rajani1, Phil Chowienczyk, Simon Redwood
1Department of Cardiology, St Thomas' Hospital, Lambeth Palace Road, London, UK.
Objectives:
To determine the relationship between brachial blood pressure, and transfer function-estimated and invasively measured central aortic pressure in patients with at least moderate symptomatic aortic stenosis.
Methods:
Fourteen patients aged 54-81 years with mean (SD) effective valve area of 0.69 (0.20) cm2, undergoing coronary angiography, had simultaneous peripheral and central aortic blood pressure measurements. Brachial blood pressure was determined by an oscillometric method. Aortic pressure was measured directly using pressure transducer tipped catheters, and estimated indirectly by the application of a transfer function to a radial arterial waveform obtained by tonometry.
Results:
Measured aortic systolic pressure did not differ significantly from brachial pressure [mean difference (SD) 2 (9) mmHg, P = not significant (NS)]. Transfer function estimates of central systolic pressure obtained from the radial waveform calibrated from brachial pressure were less accurate [mean difference -8 (7) mmHg, P = 0.001]. Recalibration of the radial waveforms using the invasive mean and diastolic blood pressure improved the agreement [mean difference -2 (6) mmHg, P = NS] but did not provide a better estimate than brachial blood pressure. The accuracy of noninvasively estimated subendocardial viability ratio was substantially improved by recalibration of radial arterial waveforms using corrected ejection time.
Conclusion:
In patients with aortic stenosis there is clinically acceptable agreement between noninvasive brachial pressure and directly measured central aortic pressure.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure