Related Experiment Video
Updated: Jun 27, 2026

Fabricating a UV-Vis and Raman Spectroscopy Immunoassay Platform
Published on: November 10, 2016
Surface characterization of immunosensor conjugated with gold nanoparticles based on cyclic voltammetry and X-ray
Lee-Jene Lai1, Yaw-Wen Yang, Yao-Kwang Lin
1National Synchrotron Radiation Research Center, 101 Hsin-Ann Road, Hsinchu Science Park, Hsinchu 30076, Taiwan, ROC. jene@nsrrc.org.tw
Abstract:
This investigation describes the surface characterization of rabbit immunoglobulin G (IgG) conjugated with gold nanoparticles. Goat anti-rabbit immunoglobulin G tagged with 5nm gold nanoparticles was applied to detect the IgG. Then, the autocatalyzed deposition of Au(3+) onto the surface of anti-IgGAu increased the surface area per gold nanoparticle. The immobilization chemistries and the atomic concentrations of Au(4f), P(2p), S(2p), C(1s), N(1s) and O(1s) of the resulting antibody-modified Au electrodes were determined by X-ray photoelectron spectroscopy (XPS). The sulfur that is involved in the cysteamine binding and the enlargement of the gold nanoparticles are identified using cyclic voltammetry. The results reveal that the surface area per gold particle, following the autocatalyzed deposition Au(3+) on the surface of anti-IgGAu, was approximately seven times higher than that before deposition.