Related Experiment Video
Updated: Jun 27, 2026

Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
Optimum frequency of office blood pressure measurement using an automated sphygmomanometer
Martin G Myers1, Miguel Valdivieso, Alexander Kiss
1Division of Cardiology, Schulich Heart Centre bInstitute for Clinical and Evaluative Sciences, Sunnybrook Health Sciences Centre cDepartment of Medicine, University of Toronto, Ontario, Canada. martin.myers@sunnybrook.ca
Objective:
To determine the optimum interval between serial blood pressure measurements using an automated BpTRU sphygmomanometer.
Methods:
Two groups of 200 patients each had automated office measurements taken using the BpTRU device at either 1-min or 2-min intervals from the start of one reading to the start of the next reading with a 24-h ambulatory blood pressure (ABP) recording being performed. Another series of 50 patients had BpTRU readings taken at 1-min and 2-min intervals before and after 24-h ABP monitoring. The difference between the mean awake ABP and the mean automated office BP readings were compared for recordings taken at 1-min versus 2-min intervals.
Results:
In the between-patient comparison (n=400), mean awake ABP was similar to automated BP recordings in the examining room at either 1-min or 2-min intervals except for a slightly lower (-4 mmHg) diastolic BP with the 1-min interval (P<0.01 vs. ABP). In the within-patient comparison (n=50), there was no consistent difference between automated BP readings taken in the examining room at 1-min versus 2-min intervals. Overall, the mean automated BP values tended to be slightly lower than the mean awake ABP.
Conclusion:
Automated measurement of BP in the office setting with devices such as the BpTRU can be taken as frequently as every 1 min without affecting the accuracy of the reading. Small differences in BP between the 1 and 2-min settings and between the automated BpTRU and ABP readings were within accepted clinical standards for validation criteria.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...