Related Experiment Video
Updated: Jun 27, 2026

Multimodal Imaging and Spectroscopy Fiber-bundle Microendoscopy Platform for Non-invasive, In Vivo Tissue Analysis
Published on: October 17, 2016
Multispectral confocal microendoscope for in vivo and in situ imaging
Houssine Makhlouf1, Arthur F Gmitro, Anthony A Tanbakuchi
1University of Arizona, Department of Radiology, P. O. Box 245067, Tucson, Arizona 85724-5067, USA.
Abstract:
We describe the design and operation of a multispectral confocal microendoscope. This fiber-based fluorescence imaging system consists of a slit-scan confocal microscope coupled to an imaging catheter that is designed to be minimally invasive and allow for cellular level imaging in vivo. The system can operate in two imaging modes. The grayscale mode of operation provides high resolution real-time in vivo images showing the intensity of fluorescent signal from the specimen. The multispectral mode of operation uses a prism as a dispersive element to collect a full multispectral image of the fluorescence emission. The instrument can switch back and forth nearly instantaneously between the two imaging modes (less than half a second). In the current configuration, the multispectral confocal microendoscope achieves 3-microm lateral resolution and 30-microm axial resolution. The system records light from 500 to 750 nm, and the minimum resolvable wavelength difference varies from 2.9 to 8.3 nm over this spectral range. Grayscale and multispectral imaging results from ex-vivo human tissues and small animal tissues are presented.
Related Concept Videos
Confocal Fluorescence Microscopy
Three-Dimensional Microscopy in Microbiology