Automated quantification of colonic crypt morphology using integrated microscopy and optical coherence tomography
Related Experiment Video
Updated: Jun 27, 2026

Quantification of Colonic Stem Cell Mutations
Published on: September 25, 2015
Xin Qi1, Yinsheng Pan, Zhilin Hu
1Case Western Reserve University, Department of Biomedical Engineering, Cleveland, Ohio 44106, USA.
Abstract:
Colonic crypt morphological patterns have shown a close correlation with histopathological diagnosis. Imaging technologies such as high-magnification chromoendoscopy and endoscopic optical coherence tomography (OCT) are capable of visualizing crypt morphology in vivo. We have imaged colonic tissue in vitro to simulate high-magnification chromoendoscopy and endoscopic OCT and demonstrate quantification of morphological features of colonic crypts using automated image analysis. 2-D microscopic images with methylene blue staining and correlated 3-D OCT volumes were segmented using marker-based watershed segmentation. 2-D and 3-D crypt morphological features were quantified. The accuracy of segmentation was validated, and measured features are in agreement with known crypt morphology. This work can enable studies to determine the clinical utility of high-magnification chromoendoscopy and endoscopic OCT, as well as studies to evaluate crypt morphology as a biomarker for colonic disease progression.