Related Experiment Video
Updated: Jun 27, 2026

Design of a Biocompatible Drug-Eluting Tracheal Stent in Mice with Laryngotracheal Stenosis
Published on: January 21, 2020
[Preparation of a biodegradable drug-eluting stent in myocardium channel]
Feng Lü1, Tian-Jun Liu, Jian Zhao
1Tianjin Key Laboratory of Biomaterial Research, Institute of Biomedical Engineering, CAMS and PUMC, Tianjin 300192, China.
Objective:
To prepare a biodegradable drug-eluting stent in myocardium channel and evaluate its effect on myocardium channel after transmyocardial revascularization (TMR).
Methods:
A biodegradable drug-eluting stent was prepared using poly (epsilon-caprolactone) (PCL), bovine serum albumin (BSA), and poly (D, L-lactide-co-glycolide) (PLGA) as material of stent, model protein drug, and drug carrier respectively. The amount of BSA in stent and in vitro released BSA of stent were determined by the Coomassie brilliant blue assay. The mechanical strength of stent was tested by universal material testing machines. The material and structure of stent was characterized by nuclear magnetic resonance spectroscopy. The effect of stent on myocardium channel after TMR was evaluated in vivo by a standard animal model of chronic myocardial ischemia in miniswine.
Results:
The stent could carry 13.1 microg BSA per mg of stent and the stent could release about 95% of BSA after 30 days. The stent diminished 80% of initial scale under the stress of 1.7 Mpa. It also kept the myocardium channel patency after TMR.
Conclusions:
A biodegradable drug-eluting stent in myocardium channel was successfully prepared. It can sustain the pressure from the heart and achieve the controlled release of drug. The stent can ensure the myocardium channel patency after TMR.