Rapid solid-phase extraction method to quantify [(11)C]-verapamil, and its [(11)C]-metabolites, in human and macaque
Jashvant D Unadkat1, Francisco Chung, Lucy Sasongko
1Department of Pharmaceutics, University of Washington, Box 357610, Seattle, WA 98195, USA. jash@u.washington.edu
Nuclear Medicine and Biology
|November 26, 2008
Summary
A new solid-phase extraction (SPE) method rapidly quantifies [(11)C]-verapamil and its metabolites in plasma. This technique is crucial for accurately measuring P-glycoprotein (P-gp) activity in PET imaging studies.
Area of Science:
- Pharmacokinetics and Drug Metabolism
- Radiochemistry and Imaging
- Biomarker Quantification
Background:
- P-glycoprotein (P-gp) is a key efflux transporter impacting drug delivery to the brain and fetus.
- Positron emission tomography (PET) ligand [(11)C]-verapamil measures in vivo P-gp activity.
- Verapamil metabolism necessitates accurate quantification for interpreting P-gp activity.
Purpose of the Study:
- Develop a rapid solid-phase extraction (SPE) method.
- Separate and quantify [(11)C]-verapamil and its radiolabeled metabolites in plasma.
- Facilitate accurate P-gp activity assessment in PET studies.
Main Methods:
- High-performance liquid chromatography (HPLC) identified [(11)C]-D-617/717 as the major metabolite.
- Sequential and differential pH elution on C(8) SPE cartridges was employed.
- Recovery was validated using nonradioactive standards and HPLC analysis.
Main Results:
- SPE method achieved >85% recovery for verapamil and D-617/717.
- Significant plasma concentrations of D-617/717 and unknown polar metabolites were detected in humans and nonhuman primates.
- SPE and HPLC methods demonstrated comparable quantification accuracy.
Conclusions:
- The SPE method enables simultaneous processing of multiple samples in under 5 minutes.
- This rapid quantification is vital given the short half-life of Carbon-11.
- The method is a valuable tool for determining [(11)C]-verapamil and metabolite concentrations in plasma.