Tympanometric experiments on retracted ear drums--does tympanometry reflect the true middle ear pressure?
Hans Wilhelm Pau1, Christoph Punke, Tino Just
1ENT Department, Medical School, University of Rostock, Rostock, Germany. hans-wilhelm.pau@med.uni-rostock.de
Conclusions:
Tympanometry in ears with retracted or partially atelectatic tympanic membranes does not reflect the true middle ear (ME) pressure. The position of the tympanogram peak depends on the size of a retraction pocket and the remaining ME gas volume. Thus tympanometry in such cases cannot be used for measurements of ME pressure.
Objective:
To prove the hypothesis that tympanometry in ears with retraction pockets or atelectasis does not measure ME pressure correctly.
Materials And Methods:
Tympanometry was performed in a simplified artificial ME model, in which different 'tympanic membranes' with pockets could be integrated.
Results:
We found a shift of the tympanogram peaks, either towards negative or positive pressures, depending on the position of the pocket: in case of an 'inward' direction (towards the ME like a retraction pocket) the peak was in the negative pressure range, in case of an 'outward' direction (bulging into the ear canal) the shift was towards positive pressures. The shift increased with decreasing volume behind the tympanic membrane. The effect cannot be simply be explained by hysteresis but by the air 'cushion' that is 'trapped' behind the membrane, limiting its excursions during the tympanometric sweep.
Related Concept Videos
Anatomy of the Ear
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Sound as Pressure Waves
The pressure fluctuation depends on the difference in displacements between the successive points in the...
Physical Assessment of the Respiratory Tract III: Percussion
Percussion in Respiratory Assessment
Percussion evaluates underlying tissue composition with audible and tactile vibrations,...