The importance of flow and pressure release in emergency jet ventilation devices
Ankie Hamaekers1, Pieter Borg, Dietmar Enk
1Department of Anesthesiology, Maastricht University Medical Centre, PO Box 5800, Maastricht 6202 AZ, The Netherlands. a.hamaekers@mumc.nl
Background:
Several self-assembled devices, consisting of a three-way stopcock connected to a high pressure oxygen source, have been proposed for transtracheal jet ventilation in an emergency situation. As a three-way stopcock acts as a 'flow splitter' it will, when connected to a continuous oxygen flow, never ensure total flow and pressure release through its side port. The aim of the present study was to measure the efficacy of flow and pressure release of three previously described self-assembled jet devices and one commercially available tool.
Methods:
In a laboratory setting simulating an obstructed upper airway the generated pressure at the cannula tip (PACT) during the expiration phase was measured in three self-assembled jet devices consisting of a three-way stopcock with an inner diameter of 2 mm (device A), 2.5 mm (device B), and 3 mm (device C), respectively, and in the Oxygen Flow Modulator (OFM) at oxygen flows of 6, 9, 12, and 15 l min(-1).
Results:
The PACT of device A at on oxygen flow of 15 l min(-1) was 71.1 (+/-0.08) cm H(2)O. At a reduced flow of 9 l min(-1) the PACT of device A was still 25.8 (+/-0.08) cm H2O. In device B and C the PACT was 35.6 (+/-0.04) and 17.6 (+/-0.04) cm H2O, respectively, at an oxygen flow of 15 l.min(-1). In contrast, the PACT in the OFM (five side holes open) was 4.4 (+/-0.02) cm H2O at the same flow.
Conclusion:
In case of complete upper airway obstruction the OFM provides sufficient flow and pressure release, whereas the self-assembled jet devices tested are inherently dangerous constructions.
Related Concept Videos
Free Jet
Design Example: Flow Through a Fire Extinguisher
The key to understanding how the...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Cardiopulmonary Resuscitation II: ACLS Airway Management