Measuring Motor Coordination in Mice
Published on: May 29, 2013
Calibration of rotational acceleration for the rotarod test of rodent motor coordination
Martin Bohlen1, Andy Cameron, Pamela Metten
1Department of Psychology, University of North Carolina at Greensboro, NC 27402-6170, USA.
Abstract:
The latency of mice and rats to fall from the accelerating rotarod can differ markedly between laboratories using the same brand of rod as well as between studies using different kinds of rods. These discrepancies can arise from different rod diameters, surface textures, test protocols, or laboratory environmental factors beyond the test itself, but it is also possible that the actual acceleration rates of the different rods do not correspond to the nominal rates set on the devices. This paper describes a simple method to measure acceleration rate of the rotarod and to set the rate to a desired value for any brand of rod.