An original calibration technique for soft x-ray detectors and its use in the Tore Supra tomographic system
1Euratom-CEA Association, DSM-DRFC, Cadarache, 13108 St Paul lez Durance, France.
Abstract:
This paper describes in detail the recent progresses which have been made in Tore Supra for developing a new technique of calibration of the soft x-ray (SXR) detectors in the range 1-30 keV. The diode response as a function of the flux of photons resulted accurately linear over almost three orders of magnitude. Apart from a limited number of deficient detectors, promptly replaced with new ones, the spread of the diode responses (84 detectors) is about 20% total of the average value. It allowed the derivation of the calibration factor for each detector of the tomography system. The effect of the environmental temperature, in the range 15 degrees-40 degrees, has also been studied, revealing that up to 35 degrees the linearity of diode response and these calibration factors remain constant. It demonstrates the capability of discriminating slight and localized changes in the two dimensional spatial distribution of the SXR intensity.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...