Estimation of copy number using SYBR Green: confounding by AT-rich DNA and by variation in amplicon length
James M Colborn1, Brian D Byrd, Ousmane A Koita
1Department of Tropical Medicine, and the Center for Infectious Diseases, Tulane University Health Sciences Center, New Orleans, Louisiana 70112, USA. fue6@cdc.hhs.gov
Abstract:
Although SYBR Green is used to estimate copy number, its fluorescence varies with amplicon length and adenine/thymine (AT) content. As a result, threshold cycle (Ct) values obtained using real-time polymerase chain reaction (PCR) are lower for longer amplicons (P<0.001) and amplicons with greater AT content (P<0.001). In contrast, neither amplicon length nor AT content affects the Ct with TaqMan probes or LUX-labeled primers. Because SYBR Green yields lower Cts with AT-rich templates and longer templates, it overestimates copy number for those templates. Therefore, sequence-specific methods such as TaqMan probes or LUX-labeled primers should be considered when using real-time PCR to estimate copy number if the amplicons generated are AT-rich or vary in length.