Use of refractometry for determination of psittacine plasma protein concentration
Carolyn Cray1, Marilyn Rodriguez, Kristopher L Arheart
1Division of Comparative Pathology, Department of Pathology, University of Miami Miller School of Medicine, FL 33101, USA. c.cray@miami.edu
Background:
Previous studies have demonstrated both poor and good correlation of total protein concentrations in various avian species using refractometry and biuret methodologies.
Objectives:
The purpose of the current study was to compare these 2 techniques of total protein determination using plasma samples from several psittacine species and to determine the effect of cholesterol and other solutes on refractometry results.
Methods:
Total protein concentration in heparinized plasma samples without visible lipemia was analyzed by refractometry and an automated biuret method on a dry reagent analyzer (Ortho 250). Cholesterol, glucose, and uric acid concentrations were measured using the same analyzer. Results were compared using Deming regression analysis, Bland-Altman bias plots, and Spearman's rank correlation.
Results:
Correlation coefficients (r) for total protein results by refractometry and biuret methods were 0.49 in African grey parrots (n=28), 0.77 in Amazon parrots (20), 0.57 in cockatiels (20), 0.73 in cockatoos (36), 0.86 in conures (20), and 0.93 in macaws (38) (P< or =.01). Cholesterol concentration, but not glucose or uric acid concentrations, was significantly correlated with total protein concentration obtained by refractometry in Amazon parrots, conures, and macaws (n=25 each, P<.05), and trended towards significance in African grey parrots and cockatoos (P=.06).
Conclusions:
Refractometry can be used to accurately measure total protein concentration in nonlipemic plasma samples from some psittacine species. Method and species-specific reference intervals should be used in the interpretation of total protein values.