Complications encountered with the use of constrained acetabular prostheses in total hip arthroplasty
Christophe Pattyn1, Roel De Haan, Alex Kloeck
1Department of Orthopedic Surgery and Traumatology, Ghent University Hospital, Gent, Belgium.
Abstract:
At our department, 46 constrained acetabular components in 38 patients were placed successively for a period of 4 years. Indications included recurrent dislocation, septic and aseptic loosening with extensive bone loss, tumor surgery with extensive bone resection, and instability due to neurologic impairment. Because 2 cup failures and 10 dislocations were observed with the constrained devices at 4 to 7 years of follow-up, the authors started to use large-diameter metal-on-metal bearings for similar indications. A series of 36 such bearings in 38 patients revealed only one cup failure and one dislocation at 2 to 4 years of follow-up. Although the 2 series are different and therefore difficult to compare, the authors recommend judicious use of constrained devices because of the high failure rate (26%) and consideration of alternative options such as the use of large-diameter metal-on-metal bearings.