Related Experiment Video
Updated: Jun 27, 2026

Source and Route of Pyrrolizidine Alkaloid Contamination in Tea Samples
Published on: September 28, 2022
Pesticide analysis in herbal infusions by solid-phase microextraction and gas chromatography with atomic emission
Natalia Campillo1, Rosa Peñalver, Manuel Hernández-Córdoba
1Department of Analytical Chemistry, Faculty of Chemistry, University of Murcia, E-30071 Murcia, Spain.
Abstract:
A direct immersion solid-phase microextraction (SPME) procedure was used in combination with capillary gas chromatography with atomic emission detection (GC-AED) for the determination of 10 pesticides (organochlorines, organophosphorus compounds and pyrethrins) in herbal and tea infusions. Ionic strength, sample dilution and time and temperature of the absorption and desorption stages were some of the parameters investigated in order to select the optimum conditions for SPME with a 100mum PDMS fiber-coating. Element-specific detection and quantification was carried out by monitoring the chlorine (479nm) and bromine (478nm) emission lines, which provided nearly specific chromatograms. Calibration was carried out by using a spiked sample infusion. The detection limits varied between 11.9ngml(-1) for deltamethrin and 0.03ngml(-1) for p,p'-DDE and p,p'-DDD. The recoveries ranged from 73.5% for deltamethrin to 108.3% for p,p'-DDT in a spiked white tea infusion. Two of the eight samples analyzed contained low levels of some the pesticides considered.