Related Experiment Video
Updated: Jun 27, 2026

Single Liposome Measurements for the Study of Proton-Pumping Membrane Enzymes Using Electrochemistry and Fluorescent Microscopy
Published on: February 21, 2019
Imaging local proton fluxes through a polycarbonate membrane by using scanning electrochemical microscopy and
1Institut für Physikalische Chemie I, Albert-Ludwigs-Universität Freiburg, Albertstr. 21,D-79104 Freiburg, Germany.
Abstract:
A new application of scanning electrochemical microscopy (SECM) to probe the transport of protons through membranes is described. Herein, a probe ultramicroelectrode (UME) is modified with a self-assembled monolayer (SAM) of 11-mercaptoundecanoic acid to qualitatively image areas within different pH regions above a track-etched membrane. The current response of the modified electrode in the presence of potassium hexacyanoferrate as electroactive component is different in acidic and alkaline solutions. Depending on the pH value of the solution, the SAM-covered electrode exposes either a neutral or a negatively charged insulating monolayer at pH 3 or 7, respectively, which leads to an increase/decrease in the faradaic current due to electrostatic interactions between the neutral/charged surface and the charged redox mediator. Therefore, local pH changes in the close vicinity of a membrane-like substrate lead to different current responses recorded at the tip electrode when scanning above the surface.