A sensitive flow analysis system for the fluorimetric determination of low levels of formaldehyde in alcoholic beverages
Related Experiment Video
Updated: Jun 27, 2026

Highly Sensitive and Rapid Fluorescence Detection with a Portable FRET Analyzer
Published on: October 1, 2016
Fabio Santos de Oliveira1, Eliane Teixeira Sousa, Jailson B de Andrade
1Instituto de Química, Universidade Federal da Bahia, Campus Universitário de Ondina, 40170-290 Salvador, BA, Brazil.
Abstract:
A sensitive FIA method was developed for the selective determination of formaldehyde in alcoholic beverages. This method is based on the reaction of Fluoral-P (4-amine-3-pentene-2-one) with formaldehyde, leading to the formation of 3,5-diacetyl-1,4-dihydrolutidine (DDL), which fluoresces at lambda(ex)=410nm and lambda(em)=510nm. The analytical parameters were optimized by the response surface method using the Box-Behnken design. The proposed flow injection system allowed for the determination of up to 3.33x10(-5)molL(-1) of formaldehyde with R.S.D.<2.5% and a detection limit of 3.1ngmL(-1). The method was successfully applied to determine formaldehyde in alcoholic beverages, without requiring any sample pretreatment, and the results agreed with the reference at a 95% confidence level by paired t-test. In the optimized condition, the FIA system proved able to analyze up to 60 samples/h.