Di-(2-ethylhexyl)phthalate leakage and color changes in endotracheal tubes after application in high-risk newborns
Giuseppe Latini1, Claudio De Felice, Antonio Del Vecchio
1Division of Neonatology, A. Perrino Hospital, Brindisi, Italy. gilatini@tin.it
Background:
Polyvinylchloride (PVC) is often adopted for making medical devices.
Objective:
Testing the hypothesis that materials degradation occurs in PVC endotracheal tubes during infant ventilation, thus releasing the known toxic plasticizer di-(2-ethylhexyl)phthalate (DEHP).
Materials And Methods:
Endotracheal tubes degradation was assessed by: (1) analysis of color and spectral changes in endotracheal tubes after use in the 400- to 700-nm range and compared to virgin samples. Color changes were expressed as euclidean distances in the Commission International de l'Eclairage Laboratory and Lightness-Chroma-Hue (Munsell's) color spaces (i.e., Delta E, Delta L, Delta a, Delta b, Delta C, and Delta H units); (2) DEHP leakage was assessed by thermal characterization by differential scanning calorimetry (DSC) and thermal gravimetric analysis (TGA). Data from used tubes were compared with those from either virgin devices or submitted to artificial aging, including O(2)-induced oxidation, washing, UV photodegradation and exposition to acid solutions.
Results:
Significant color differences of 13.33 +/- 0.85 Delta E were evidenced in used endotracheal tubes, as compared to virgin samples (p < 0.0001). Chromatic changes were independent from intubation duration. DSC and TGA analyses showed a loss of DEHP from used tubes, as indicated by increase of the glass transition temperature (Tg) and DEHP weight loss. In vivo application of tubes was associated with an aging process whose effects on the optical and spectral properties were similar to those observed following artificial aging processes.
Conclusions:
Our findings indicate that significant spectrocolorimetric, DSC and TGA changes occur in endotracheal tubes after application, thus demonstrating for the first time the occurrence of in vivo materials degradation and DEHP leakage from medical devices worldwide used for neonatal ventilation.
More Related Videos
06:59Intrathecal Application of a Fluorescent Dye for the Identification of Cerebrospinal Fluid Leaks in Cochlear Malformation
Published on: February 29, 2020
06:47Effects of Exposure of Formaldehyde to a Rat Model of Atopic Dermatitis Induced by Neonatal Capsaicin Treatment
Published on: September 27, 2017
Related Concept Videos
Pharmacokinetics in Pediatric Patients: Drug Distribution
Pharmacokinetics in Pediatric Patients: Overview and Drug Absorption
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...