Preventing central venous catheter-associated bloodstream infections: development of an antiseptic barrier cap for
Steve Z Menyhay1, Dennis G Maki
1Menyhay Healthcare Systems LLC, Fresno, CA, USA. menyhay@psnw.com
Background:
Reports of outbreaks associated with the use of needle-free valve connectors suggest 2 common risk factors: (1) poor adherence to disinfection practices before use and (2) a design that allows contamination when not in use. Swabbing a membranous septum with 70% isopropyl alcohol may not eliminate septal surface contamination. Frequent access through and handling of needle-free connectors also puts patients at increased risk of central venous catheter-related bloodstream infections (CR-BSIs). A novel antiseptic barrier cap has been designed to maintain health care worker safety and eliminate the vulnerabilities of existing systems to contamination and CR-BSIs for patients at risk.
Methods:
A prospective in vitro study compared the effectiveness of standard disinfection of needleless luer-activated valve connectors with 70% isopropyl alcohol and the effectiveness of an antiseptic barrier cap that, when threaded onto a luer-activated connector, rapidly sterilizes a heavily contaminated surface. Standard disinfection was done by 3- to 5-second swabbing using a sterile commercial pledget of 70% isopropyl alcohol. The antiseptic barrier cap comprises an outer cap with internal female threads and a spike inside the closed end, a capsule containing 0.25 mL of 2% chlorhexidine gluconate in 70% isopropyl alcohol, and a sponge between the septum and the capsule. When the cap is threaded onto a luer-adaptable needleless connector, the spike ruptures the capsule, saturating the sponge with the antiseptic.
Results:
All 15 (100%) of the precontaminated positive control connectors not disinfected before entry showed transmission of Enterococcus faecalis across the membranous septum (4500-28,000 colony-forming units), and 20 (67%) of 30 connectors disinfected with 70% alcohol showed transmission (442-25,000 colony-forming units). Of 60 needle-free connectors disinfected with the antiseptic barrier cap, 1 (1.6%) showed transmission (P < .001).
Conclusion:
An antiseptic barrier cap was highly effective in sterilizing the septum of a needle-free valve connector and preventing entry of any microorganisms, even with heavy contamination of the septum. This new technology should now be evaluated in a clinical trial with CR-BSI as the primary outcome measure.
Related Concept Videos
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Hand hygiene
Hand washing...
Endocarditis IV: Nursing Management
Asepsis